My result’s color is white, what could the issue be?

If you get white colors after a test, this might be due to the lack of water in the testing chamber and/or lack of chemical reagents.

What do pale test result colors mean?

If you get pale colors while testing, this indicates a couple of issues:

  1. Chemical reagents need to be re-filled in the vial
  2. Lack of magnetic stirrers in the vials and/or testing chambers
  3. Syringe needs replacing
  4. The testing chamber needs cleaning

If my colors while testing are wrong (not as indicated on the test kit), what do I need to do?

  1. Pumps need to be re-calibrated
  2. Lack of magnetic stirrers in the vials could be the issue
  3. Chemical reagents might be contaminated or chemical positions are wrong: Make sure the RO water is filling the RO chamber. If not, check if the silicon “RODI” tube is connected to the RO tank with no air bubbles
  4. Calibrate your RODI chamber pump (check calibration procedure).

My LED indicators are consistently red, what does that mean?

If your LED indicators is consistently red, kindly fill out the customer support form indicating the issue and our team will help with the troubleshooting.

I am getting a notification on the Reef Kinetics App that my waste is full and I cannot perform a test, what do I do?

Log in to the App Go to the Bot and click on Maintenance Click on “Empty” next to Waste You will now be able to perform a test

How do I replace my SD Card if it is corrupted?

Do not change SD unless you contact our customer support team first. Make sure your ReefBot is turned off. Unscrew the back Acrylic of the ReefBot. Remo…

I am getting a message on the Reef Kinetics App that my Syringe needs replacement and I cannot perform a test, what do I do?

  1. Log in to the App
  2. Go to the Bot and click on Maintenance
  3. Click on Replace next to Syringe
  4. You will now be able to perform a test

When do I clean the reaction and rinsing chambers and how?

Clean the reaction and rinsing chambers every 100 tests or whenever needed by following the procedure below:

• Soak both chambers with warm (not hot) water for about two minutes.
• Scrub firmly with tissue paper or a small brush.

When do I replace the syringe and needle on my ReefBot?

You need to change them every 50 tests or whenever needed.

When do I replace the pump heads?

You need to replace them every 6 months or at any sign of malfunctioning.

How can I calibrate RO and reaction chambers?

Open the Reef Kinetics App. Select the Bot that is assigned to your tank from the menu tab. Go to maintenance page and follow the below points: 1. Select t…

How full do I need to fill my vial with chemicals?

Your vial should be minimum 55% full in order for you to perform a test.

In this case, please make sure your vial is always full. You will get notified when your vial is about to get below the 55%.